Catalog Number 1025DS
Delivery Rate - 2 to 10 LPM
Maximum Recommended Flow (@ nominal outlet pressures of zero and 7 kPa)** - 10 LPM
Outlet Pressure - 20.0 ± 1.0 psi (138 kPa +/- 7 kPa)
Auxiliary Oxygen Port - Outlet Pressure: <15 psi Outlet Flow: 2 LPM
Electrical Rating -120 V, 60 Hz, 6.1 Amp
Operating Voltage Range - 102-132 V~, 60 Hz
Oxygen Percentage - 2-10 LPM=87%-96%
Operating Altitude (tested at 70°F {21°C} only) 0-1500 M (0-5000 ft) - Across the voltage range: No degradation of performance
Operating Environment Range* 41°F (5°C) to 95°F (35°C), humidity range of 15% to 93% non-condensing - No degradation in performance across the operating voltage range. Power Consumption -120 vac, 60HZ: 639 watts average
Weight - 42 lbs. (19 Kilograms)
Safe Working Load - 53 lbs. (24 Kilograms)
Sound Pressure Level at 3 and 10LPM - <59 dBA
Dimensions - 24.5”H x 13.5”W x 12”D (62.2 x 34.2 x 30.4 cm)
Maximum Limited Pressure - 35 PSIG (241 kPa)
Operating System - Time Cycle / Pressure Swing
Low Oxygen Indicator - <82% low oxygen <60% very low oxygen
Storage Conditions - -13°F (-25°C) to 158°F (70°C), humidity range of 15% to 93% non-condensing