Indications for use -
-
Life support for patients over 5 kg*
Circuit -
-
Single circuit with leak
-
Single circuit with valve
Therapy -
-
Valve circuit therapy: CPAP, ACV, PACV, P-SIMV, V-SIMV, PS
-
Leak circuit therapy: CPAP, ST, PAC, iVAPS*
-
Manual breath
-
Sigh breath (recruitment)
-
Apnoea ventilation- Volume and pressure
-
Preset programs - 2
Pressure Range and Settings -
-
Delivered pressure (cm H2O) - : 2–50 (leak circuit) , 0–50 (valve circuit)
-
Delivered tidal volume (mL) Adult: 100–2500 mL
-
Paediatric: 50–300 mL
-
Respiratory rate Adult: Off, 2–50 bpm
-
Paediatric: Off, 5–80 bpm
Data -
-
On screen 30 days of summary data
-
Real-time pressure and flow (25 Hz)
-
ResScan (via download) 365 days of summary data
-
7 days of detailed data
Alarms - a wide range of settable alarms are available including FiO2 and SpO2, when connected.
Dimensions-
-
Weight - 3.2 kg
-
Dimensions (L x W x H) - 285 mm x 215 mm x 93 mm
Sound power level 43 ± 3 dBA**
Sound pressure level 35 ± 3 dBA**
* iVAPS therapy mode is indicated for patients weighing 30 kg and above